Radiation and magnetic shielding contractor in Saudi Arabia
We build the protective envelope around diagnostic and therapy rooms: lead lining for ionising radiation, and RF (Faraday cage) and magnetic shielding for MRI. Every layer is built to the OEM site-planning guide and the radiation protection physicist’s report.

Two different problems, two different shields
Radiation shielding contractors build the rooms that keep an imaging scanner’s energy where it belongs. In CT, PET-CT and X-ray rooms the hazard is ionising radiation, which is stopped by dense material, usually lead; a radiation protection physicist calculates how much each wall, floor, ceiling, door and window needs and records it in a shielding report. In an MRI suite the problem is different: the room is shielded from radio-frequency interference by a continuous conductive envelope, usually copper, aluminium or galvanised steel, and sometimes the magnet’s fringe field is contained with steel plate, as the manufacturer’s site-planning guide specifies. EGC builds both kinds of room from those two documents. We detail every junction, door, window and penetration on shop drawings before work starts, install the shielding with our own crews, and support the RF attenuation test or radiation survey before the room is released to the equipment manufacturer’s installation team.
The weak points are always the same: doors, windows, junctions and every pipe, duct and cable that crosses the shield. That is why we coordinate shielding with the door and healthcare MEP packages instead of treating it as a separate trade.
Where it is used
- MRI suites
- CT rooms
- PET-CT suites
- X-ray and fluoroscopy rooms
- Nuclear medicine rooms
- Radiation therapy vaults
What we deliver
- MRI suites with RF (Faraday cage) and magnetic shielding
- Lead-lined walls, ceilings and floors for CT, PET-CT and X-ray rooms
- Nuclear medicine rooms and radiation therapy vaults
- Lead-glass viewing windows and RF windows
- Continuous-lining details at junctions, boxes and penetrations
- Retrofit shielding for existing diagnostic rooms
From our sites

Lead-lined room under construction, with metal studs and a ladder 
Lead-lined wall with a steel frame and a service opening 
Sliding door with radiation warning signs and a card reader beside it
MRI and CT shielding compared
| MRI suite | CT, PET-CT and X-ray rooms | |
|---|---|---|
| What it controls | Radio-frequency interference getting in, and the magnet’s fringe field reaching outside | Ionising radiation (X-rays, or the 511 keV photons from PET tracers) escaping the room |
| Typical material | A continuous conductive envelope of copper, aluminium or galvanised steel, plus steel plate where the manufacturer specifies it | Dense material, usually lead sheet or lead-lined board |
| Who sets the amount | The manufacturer’s site-planning guide | The radiation protection physicist’s shielding report |
| Usual weak points | Door, window, waveguides and filters, any unsealed seam | Door, viewing window, junctions, boxes and penetrations |
| How it is checked | RF attenuation test | Radiation survey once the room is built |
Who specifies what, and who builds it
| Who | What they do |
|---|---|
| Equipment manufacturer | Publishes the site-planning guide for the exact model: room size, RF and fringe-field limits, services and penetrations. |
| Radiation protection physicist | Calculates the shielding each surface needs and issues the shielding report. |
| Hospital design team and main contractor | Provide the structure, the layout and the programme the shielding has to fit into. |
| EGC | Reads both documents together, builds and inspects the shielding, and supports RF and radiation testing before handover. |
How we deliver it
Review the specification
We read the OEM site-planning guide and the physicist’s shielding report side by side and resolve conflicts before drawings are issued.
Detail the envelope
Shop drawings show every lining layer, seam, junction and penetration, so nothing is improvised on site.
Build and inspect
Our own crews install the shielding, and continuity is checked before finishes close it in.
Test and hand over
We support the RF attenuation test or the radiation survey, then release the room to the OEM installation team.
Related disciplines
Medical doors & access
Lead-lined and RF-shielded doors, hermetic sliding doors and access control — fabricated and fitted to seal controlled clinical areas.
Healthcare MEP
Mechanical, electrical and plumbing works engineered for imaging equipment — medical gases, earthing, EMI-aware routing and controlled-environment HVAC.
Infection-control surfaces & joinery
Seamless, non-porous Corian solid surfaces and medical-grade joinery from our own factory, made for clinical cleaning protocols.
Planning an MRI, CT or PET-CT room?
Send us the equipment model and the site, and we will tell you what the room needs and how we would deliver it.